Properties:
Related Categories | Additional Standards, Analytical Standards, Analytical/Chromatography, Proficiency Testing, Water Supply/Drinking Water Organics, |
packaging | ampule of 1 × 2 mL |
mfr. no. | RTC, PEO-011MA |
concentration | in acetone |
storage temp. | 2-8°C |
Description:
General description
This proficiency testing sample is produced in accordance with ISO/IEC 17043:2010. All information regarding the use of this material can be found in the reporting packet supplied for each sample.
Other Notes
Proficiency testing samples are presented as single blind. No values for the samples are released unti the testing period concludes. Values of analytes vary lot to lot.
Proficiency Testing Schedule (2017-2018)
Wastewater
WP17-1 January 18 - March 3, 2017
WP17-2 March 29 - May 12, 2017
WP17-3A May 17 - June 30, 2017
WP17-3B July 12 - August 25, 2017
WP17-4 September 27 - November 10, 2017
WP17-5 November 22 - January 5, 2018
WP18-1 January 17 - March 2, 2018
WP18-2 March 28 - May 11, 2018
WP18-3A May 16 - June 29, 2018
WP18-3B July 11 - August 24, 2018
WP18-4 September 26 - November 9, 2018
WP18-5 November 21 - January 4, 2019
Quick-Turn Study - (non-scheduled proficiency test that is available at any time)
Preparation Note
This ampule contains a concentrate of EPH/PAHs.
This PT sample is packaged as a 1.5 mL concentrate.
The solvent for this sample is acetone.
The PT sample is not preserved.
Sample preparation
To prepare solution, fill a 1000-mL volumetric flask to volume with carbon-filtered reagent-grade water at room temperature.
Open ampules of PEO-011MA (be careful to avoid cuts on glass) and immediately withdraw 1 mL.
Quickly inject the 1 mL aliquot into the reagent water in the expanded area of the volumetric flask.
Mix the sample by inverting the flask three times only and analyze immediately.
Analytes are reported in μg/L.