eCl@ss 32110501
Properties:
Related Categories | Analytical/Chromatography, HPLC, HPLC & FPLC Columns for Biopolymer Separations, TSKgel Gel Permeation Columns for Organic-Soluble Polymers, TSKgel HXL Type Gel Permeation Columns, |
mfr. no. | Tosoh, 07113 |
parameter | 1.2 mL/min flow rate |
60 °C temperature | |
application(s) | HPLC: suitable |
L × I.D. | 4 cm × 6 mm |
matrix | polymer particle platform |
matrix active group |
polystyrene-divinylbenzen |
Description:
Compatibility
THF can be replaced with benzene, chloroform, toluene, xylene, dichloromethane, dichloroethane
Physical form
Shipped in THF.
General description
The TSKgel® HXL column line consists of eight columns with different pore sizes, TSKgel G1000HXL through TSKgel G7000HXL, and three columns with an extended linear range of the calibration curve, TSKgel GMHXL, TSKgel GMHXL-L and TSKgel MultiporeHXL-M. The calibration curves are shown to the right. Main characteristics are: ultra-low sample adsorption, i.e., the columns show true size exclusion behavior for most polymers, maximum operating temperature for TSKgel HXL-type columns is 60°C for G1000HXL, G2000HXL, G2500HXL and G3000HXL, and 80°C for the remaining columns in the product line. See TSKgel HHR HT columns for operating temperatures of 140°C, the columns are shipped in THF, which can be exchanged for a limited number of organic solvents. TSKgel HXL-type columns are conventional GPC columns with column dimensions of 7.8mm ID x 30cm.
Recommended products
Discover LiChropur reagents ideal for HPLC or LC-MS analysis
Legal Information